| Ultrapure water | 0.055 µS/cm at 25°C | 0.00–2.00 µS/cm | 0.001 µS/cm or better | Approximately 0.01 cm−1 | Verifying final water quality in high-purity laboratory and analytical applications | Use temperature compensation, a clean flow-through cell, and a meter designed for very low conductivity. Carbon dioxide absorption can quickly increase the reading. |
| Distilled water | Approximately 0.5–3 µS/cm | 0–20 µS/cm | 0.01 µS/cm | Approximately 0.1 cm−1 | Checking distillation performance and laboratory water preparation | Minimize exposure to air and use a low-range conductivity cell to avoid loss of sensitivity. |
| Deionized water | Approximately 0.1–10 µS/cm | 0–20 or 0–200 µS/cm | 0.01–0.1 µS/cm | Approximately 0.1 cm−1 | Monitoring ion-exchange systems, polishing cartridges, and laboratory water outlets | Select a stable low-conductivity range and confirm that the probe material is compatible with high-purity water. |
| Reverse-osmosis permeate | Approximately 5–50 µS/cm | 0–200 µS/cm | 0.1 µS/cm | Approximately 0.1–1.0 cm−1 | Evaluating membrane performance and detecting changes in permeate quality | A wider low-range meter is practical for routine plant monitoring; automatic temperature compensation is strongly recommended. |
| Drinking water | Approximately 50–800 µS/cm | 0–2,000 µS/cm | 1 µS/cm | Approximately 1.0 cm−1 | Routine water-quality screening and process-water checks | Choose a durable probe, a readable field display, and calibration options suitable for mid-range conductivity standards. |
| Fresh surface water | Approximately 100–1,500 µS/cm | 0–2,000 or 0–20,000 µS/cm | 1–10 µS/cm | Approximately 1.0 cm−1 | Rivers, lakes, reservoirs, environmental surveys, and source-water comparison | Use temperature compensation and a probe that can be rinsed easily. Consider a protective sensor guard for field sampling. |
| Hydroponic nutrient solution | Approximately 800–3,000 µS/cm | 0–5,000 µS/cm | 10 µS/cm | Approximately 1.0 cm−1 | Managing dissolved nutrient concentration in crop-growing systems | Choose a probe resistant to fouling and a meter with automatic temperature compensation because solution temperature can vary during operation. |
| Brackish water | Approximately 1–10 mS/cm | 0–20 mS/cm | 0.01 mS/cm | Approximately 1.0–10 cm−1 | Estuarine monitoring, aquaculture, and salinity-change screening | Use a corrosion-resistant probe and verify that the meter can display both µS/cm and mS/cm. |
| Seawater | Approximately 45–60 mS/cm | 0–100 mS/cm | 0.1 mS/cm | Approximately 10 cm−1 | Marine-water testing, aquaculture, and salinity estimation | A meter limited to 12.88 mS/cm is not sufficient for seawater. Select a high-range instrument and rinse the sensor after use. |
| Calibration reference solution | 12.88 mS/cm at 25°C | At least 0–20 mS/cm | 0.01 mS/cm | Approximately 1.0 cm−1 | Calibrating and verifying meters used for medium-to-high conductivity measurements | Use a certified or traceable standard, observe its stated reference temperature, and replace contaminated or expired solution. |